EtO Sterilization: Microbiological Aspects of Process Validation
Deliberate decision making during the structuring of microbial challenges,
product loads, and biological indicators can provide a validation process for
EtO sterilization that ensures accuracy, the absence of microbes, and a smooth
testing procedure.
Susan Edel Satter and Paul J. Sordellini
A companion to
this article, EtO Sterilization: Principles of Process
Design, discussed the components of each phase of two 100% EtO with nitrogen
processes, focusing on the engineering aspects of designing EtO cycles.1
This article focuses on various approaches to medical device sterilization cycle
validation from a microbiological standpoint. The discussion assumes that the
first two stages of the validation—the commissioning and the physical
performance qualification of the sterilization chamber—have been successfully
completed. Therefore, this discussion deals solely with the third stage of EtO
validation: the microbiological performance qualification (MPQ). The topics that
are discussed include how to choose the appropriate microbial challenge for an
EtO sterilization process, approach to developing the EtO cycle, product load,
placement of biological indicators in/on the product load, method for
determining cycle lethality, and calculations to determine the D-value.
Information is also given on the documentation for the report on validation
and certification of the process, and revalidation is discussed briefly. Some
suggestions exceed the current requirements presented in international
standards, but they can enhance a validation process, resulting in a more
thorough and accurate study.
International standards ISO 9001 and 9002 present the quality system
requirements for the design, development, production, installation, and
servicing of healthcare products. The ISO 9000 series treats medical device
sterilization as a special manufacturing process because the results cannot be
verified by inspecting and testing 100% of the product at the conclusion of the
cycle. Sterilization processes must, therefore, be assessed for special
considerations, validated prior to use (or during use in certain situations),
and routinely monitored.
To design an effective validation and routine control program for a
sterilization process, the bioburden on the product and packaging must be
considered. Bioburden is defined as "the population of viable
microorganisms on a product and/or a package" and is characterized in terms
of number, identity, and resistance. A validated test method must demonstrate
that it can consistently and adequately remove the bioburden from the product
and packaging. There are various bioburden test methods and associated
validation procedures from which to choose.2
Many factors can contribute to the bioburden on product and packaging,
including the origin of the raw materials and components, transit and storage
conditions, and the manufacturing environment in which the finished products are
assembled and packaged. European standards place particular emphasis on
controlling the processes used to manufacture sterile products.3
Sterility is defined as the "state of being free from viable
microorganisms."4 Microbial death relating to the gaseous
sterilization of healthcare products is an exponential function typically
defined as the probability of a nonsterile item existing per a given number of
units in a batch. This probability, called the sterility assurance level (SAL),
defines the "probability of a viable microorganism being present on a
product unit after sterilization."4 While sterilization can
reduce the bioburden on a given product to a very low number, that probability
can never be reduced to zero. Therefore, in order to achieve the desired
bioburden levels, it is critical to design a validation program that provides a
high degree of confidence for consistent sterilization.
DETERMINING
THE APPROPRIATE MICROBIAL CHALLENGE
A biological indicator (BI) is an "inoculated carrier contained within
its primary pack providing a known resistance to the relevant process."4
There are many different types of BIs but the most common include:
- Commercial units supplied with the manufacturer's certification, such as
bacterial spore strips individually packaged in glassine envelopes or pieces
of filter paper impregnated with a certified population of a challenge
organism that has a known resistance to EtO, such as the spores of Bacillus
subtilis var. niger.
- A liquid bacterial spore suspension commercially sold and certified by the
manufacturer. The suspension is placed on or in the product, which is then
referred to as inoculated product.
- A liquid bacterial spore suspension made by a device manufacturer using a
commercial, certified strain of a bacterium.
- Strains of microorganisms that have been isolated from a device
manufacturing facility. These strains represent the most resistant organisms
found in or on the devices and are typically used for the combined BI/bioburden
method of validation.
It is crucial to ensure that the type of BI used to validate or routinely
monitor a given sterilization process is the most appropriate indicator for that
process. In addition to identity, quantitation, resistance, storage, general
directions for use, and disposal conditions, the manufacturers of BIs are
required to provide information regarding the optimal culturing conditions, such
as temperature and type of growth media.
Irrespective of which BI is chosen, the methods used to recover the challenge
organism must be validated. This recovery is expressed in terms of the percent
recovery of the original inoculum. These recovery studies can be especially
challenging when using liquid spore suspensions because of the interaction
between the suspension and the material onto which it is inoculated. The
material substrate can alter the resistance characteristics of the inoculum
because of such anomalies as spore clumping or the physical sheltering of spores
in certain sites within the product.
The goal is to kill the microbes, which means disabling their ability to
reproduce even in their most favorable growth conditions as described by their
manufacturer. There have been records of seemingly killed microbes that
regenerated when conditions become favorable.5 The user must validate
that the incubation time under the prescribed conditions is sufficient to
recover delayed growth of the organisms after exposure to a given sterilization
process. For routine processing, this time period is typically 7 days unless
validated for a shorter time period in accordance with current national
requirements.6 In such cases, periodic checks should be run to
confirm that the shorter time period yields equivalent recoveries to those
obtained from the longer incubation period. It is also important to ensure that
the incubation time is sufficient to recover growth from injured organisms
exposed to sublethal cycles. In some cases, this may mean using a 14-day
incubation period. This incubation period is also required by the U.S.
Pharmacopeia for product sterility testing.7
Just as the BI must provide a defined resistance to a specified process, it
is necessary to prove that the inherent bioburden on the product does not have a
greater resistance than the BI. Characterizing bioburden involves quantitation,
identity, and resistance of the bioburden. Several methods can be employed to
determine which BI is appropriate for a specific situation.
- Challenge organisms that make up BIs typically resist a particular
sterilization process far more than do common bioburden organisms. The
resistance of the bioburden cannot be adequately evaluated by quantitation
only, yet it must be determined if the products to be processed might
contain organisms that are more resistant than the challenge organism. After
the microbial identifications and quantitation of the bioburden have been
completed and analyzed, comparative resistance determinations of the most
resistant bioburden component or product must be calculated. A literature
review is also required.
- When microbial identifications of the bioburden are not performed, at
least one sublethal cycle should be run to compare the relative inactivation
rates of the bioburden with that of the challenge organism. Product
sterility testing, after exposure to at least one sublethal cycle under
appropriate experimental conditions, can ensure that the product's bioburden
is not more resistant than the challenge organism.
- If the quantity, identity, and resistance of the product's bioburden are
known, it might be possible to validate and routinely monitor the
sterilization process by combining BI and bioburden methods. It must be
demonstrated that the BI's degree of challenge to the sterilization process
is adequate to ensure that the process will attain the desired SAL for the
bioburden. Combining the BI and bioburden methods to determine the
appropriateness of the BI can be time-consuming and result in additional
testing costs. However, the required sterilization parameters can be more
accurately determined, which can result in reduced processing time and
reduced exposure to the sterilant.
BIs can be configured in many different ways depending on the cycle
development method chosen.
Inoculated Product. The actual product, configured and packaged as it
is intended to be sold, can be inoculated with spores of a microorganism such as
Bacillus subtilis var. niger. Direct inoculation usually uses
spores suspended in liquid, then placed on the product and dried. The product's
surface characteristics will affect the distribution of spores and may lead to a
difference in resistance behavior compared with other challenge systems.4
It is, therefore, important to achieve an even distribution of spores on the
product's surface. Indirect inoculation involves placement of a carrier, such as
filter paper that has been impregnated with spores, in the product or its
package.
Inoculated Unit. A carrier, such as a filter paper strip or disk, can
be inoculated with a population of a resistant organism, such as Bacillus
subtilis var. niger, that has been extensively characterized and
certified by the manufacturer. The resistance of this inoculated carrier must be
compared to that of the inherent bioburden of the product being validated or the
equivalent simulated product. An inoculated unit is usually used when there is
the potential that the bioburden on the product is more resistant than the
indicator organism and is required for the combined bioburden and BI cycle
development method.
Inoculated Simulated Product. A simulated product that comprises the
most difficult to sterilize portions of a device or that configuratively
represents a device family can also be directly or indirectly inoculated. This
simulated product must present the greatest challenge to the process in order to
be considered an adequate microbial challenge. Each unit must contain a
certified inoculum either in liquid form or on a carrier.
Natural Product. The inherent bioburden on the product can also be
used as the microbial challenge during validation and for routine monitoring
when the absolute bioburden method is employed for cycle development (see "EtO
Cycle Development Approaches", below).
All validation methods for EtO sterilization require that the BI used for
validation and to monitor routine cycles must be more resistant than the
bioburden of the product and be placed in a location that is more difficult to
sterilize. Comparative resistance testing is an effective means of selecting the
BI and its location in the product that presents the greatest challenge to the
sterilization process. Such an assessment should be made prior to validation as
part of determining the appropriateness of the BI. These studies are usually
carried out in small chambers that are capable of delivering rapid ramp rates,
e.g., the times required to achieve specific pressure set points.
Products should be exposed to cycles in which the only variable is the gas
exposure time period. The data obtained from this testing can be used to justify
the choice of a specific actual or simulated product to inoculate and use for
the BI. If the design of the product is such that a BI unit cannot be placed in
the part of the device that is the most difficult to sterilize, the product
should be inoculated with a liquid spore suspension to provide a known number of
viable spores. The spore suspension, materials, and techniques used should
comply with ISO 11138, parts 1 and 2.8,9
Many device manufacturers include an additional objective in their validation
plan that involves the use of external BI monitoring systems. Often referred to
as process challenge devices (PCDs), they assess the lethality of the EtO
process after the cycle has been designed. The PCDs are geometrically
distributed around the load rather than in internal locations in the case
cartons. Direct comparisons can then be made between the sterility test data
obtained from these external PCDs and the BIs placed in internal locations. A
PCD must be shown through comparative resistance studies to provide more of a
challenge to the process when it is placed in external locations in the load
than do the the BIs placed in internal locations. They usually, therefore, bear
no resemblance to the product. Examples of external PCDs are spore strips
double-packaged in plastic bags, in sealed plastic tubing, or in syringes. There
are also commercially available PCDs that are sold as ready-to-use packaged
systems. It is advisable during the validation studies to evaluate different PCD
configurations during the comparative resistance studies to determine the best
candidate. To monitor routine sterilization cycles, it must be shown at the time
of validation that the PCDs in the external locations comply with the same
requirements for resistance to sterilization.
ETO CYCLE DEVELOPMENT APPROACHES
There are three basic approaches to developing EtO sterilization cycles—the
overkill method, the combined bioburden and BI method, and the absolute
bioburden method.
The overkill method is probably the most widely used because it is relatively
easy to use and it results in a robust SAL. The method ensures that the
sterilization process will inactivate a specific number of microorganism spores
known to be resistant to the EtO sterilization process. The organism most
commonly used to monitor the overkill process is Bacillus subtilis var.
niger. A certified preparation consisting of a stated population of Bacillus
subtilis var. niger spores is inactivated through exposure to
specific cycle parameters that have been assessed to be significantly higher
than those required to kill the inherent bioburden on the product. The
parameters are increased on a routine basis to provide the desired SAL (see
"Methods for Determining Cycle Lethality," below).
The combined bioburden and BI method is used when the two are equally
resistant. This method requires routine bioburden and BI testing in addition to
a considerable amount of routine sterility testing to develop a cycle that will
inactivate the BI challenge population. The BI must be sufficiently resistant to
ensure that the EtO process will deliver the desired SAL relative to the
bioburden on the product.
The absolute bioburden method is used less frequently in cycle development
because it requires extensive testing in both the development phase and routine
processing. However, it must be used when the product's bioburden is more
resistant than the BI. Such bioburden resistance to the EtO process can be
caused by any number of factors, such as the configuration of the product, the
quantity or location of the microorganisms, or the bioburden's intrinsic
resistance. Since the bioburden on the product constitutes the essential
microbial challenge for the process, the bioburden test method must be validated
and strictly controlled. The resistant microorganisms are screened through
bioburden testing and may be isolated and propagated for use in cycle
development studies. One negative of this method is that the microorganisms'
resistance can change as a result of how they are cultured, which can adversely
affect the results of the cycle development studies. The absolute bioburden
method also requires extensive controls of the manufacturing environment in
addition to routine product bioburden monitoring and resistance studies.
DETERMINING THE APPROPRIATE PRODUCT LOAD CHALLENGE
Microbiological performance qualification (MPQ) should be performed using
specified products and packaging configured in the same manner in which they
will be routinely sterilized. For the cycle to be accurate, the product load
must represent the greatest challenge intended for future routine sterilization.
If a device manufacturer intends to use multiple load configurations on an
ongoing basis, the densest configuration should be used for the MPQ.
Each type of configuration must be documented in terms of the number of
product units per case, the number of cases per pallet, the stacking patterns on
the pallet, and the density. This documentation should be included with the
validation data. Some testing should also be conducted on the least dense
configuration, which, theoretically, presents less of a challenge to the
process. This testing can be as simple as placing thermocouples throughout the
least dense load on a routine cycle and comparing the temperature distribution
with that of the densest load. In other cases, additional microbial challenge
studies might be required. Changes in the product load must be evaluated
carefully because seemingly innocuous changes, such as changing the shrink wrap
or corrugate on the load, can have a significant effect on the cycle's efficacy
from the perspective of product sterilization.
BI PLACEMENT IN THE PRODUCT LOAD
After the product load challenge has been identified, the BI positioning and
placement can be determined. BIs should be distributed throughout the product
load and, as much as possible, in the same orientation (e.g., vertical). The
placement must include those locations that are considered to present the
greatest challenge to the process and can be the same as those used for
temperature monitoring. The ANSI/AAMI/ISO 11135-1994 standard suggests placing
two BIs at each location with a temperature-monitoring device in order to obtain
additional information on process efficacy. It also provides the following
recommendation for the number of BIs to be included in each validation cycle:
- At least 20 BIs for usable chamber volumes up to 5 m3.
- Increase the number of BIs by two for every additional 1 m3 of
usable sterilizer chamber volume between 5 and 10 m3.
- Increase the number of BIs by two more for every additional 2 m3
of usable sterilizer chamber volume above 10 m3.
The AAMI technical information report "Contract Sterilization for
Ethylene Oxide" can provide additional information on the number of BIs and
monitoring devices recommended based on product load volume.10
METHODS FOR DETERMINING CYCLE LETHALITY
Results obtained from commissioning and physical performance qualification
and monitoring devices should be used to identify critical features of the
equipment or process that can be investigated during the MPQ. For example, it is
critical that the sterilant injection time is consistent among the MPQ cycles to
ensure a uniform delivery from one cycle to the next. Even minor changes in the
sterilant injection time can result in significant differences in lethality.
The MPQ should be performed in the industrial chamber that will also be used
for routine processing unless equivalency can be demonstrated between the
industrial chamber and whatever chamber is used for the qualification.
Maintaining the precise and consistent delivery of the sterilization cycle
parameters is more difficult to accomplish in large industrial chambers than in
small test chambers. It is also important to conduct these studies using the
actual product load intended for routine sterilization. Hence, these studies are
usually conducted in large industrial chambers rather than in small test
chambers.
The MPQ can be performed by determining the lethality of the cycle on the
basis of the number of D-values applied. The D-value is defined as
"the time required to reduce a specific microbial population by 90% or one
logarithm."4 The survivor curve construction or
fraction-negative methods (described below) may be used as outlined in current
standards.4 Another means of evaluating the MPQ is the half-cycle
method, based on the number of times required to completely inactivate the BI
microorganisms with an added margin of safety. The ultimate objective of each
method is to determine the full cycle to which the product load must be exposed.
Survivor Curve Construction Method. The
survivor curve construction method involves the direct enumeration of survivors
in terms of colony forming units (CFUs) recovered after exposure to graded
amounts of the sterilization cycle. A CFU is defined as "a visible
outgrowth of a population of organisms arising from a single or multiple
cells."2 A minimum of five cycles should be run, each using
different graded time exposures to EtO.4
The parameters used, with the exception of the gas exposure time, must be kept
consistent. The first cycle is a time zero study in which the initial CFU
survival count of the BI is determined by exposing the BIs to all stages of the
process, including preconditioning if used, prior to the EtO injection phase of
the cycle. All BIs should survive because they will not be exposed to the
sterilant.
After each of the four or more additional cycles, all employing different gas
exposure time periods, the number of BIs that survive the processes are counted.
The BIs should be removed from the chamber and the load as soon as possible
within the confines of worker exposure policies. The BIs should also be tested
as soon as possible after being removed to reduce their exposure to EtO
residuals, which can affect BI survival rates. In all cases, the time intervals
between when the load is removed from the chamber, when the BIs are removed from
the load, and when they are subjected to the enumeration process must be
consistent among the cycles. Ideally, a final enumeration of the BIs from one
cycle should be obtained before the next cycle is initiated to more accurately
assess the exposure time to use. This is not always feasible because the
enumeration process can take days or weeks to complete.
The number of BIs used in each of the cycles in the study should be
statistically significant to ensure obtaining dependable data. The number can
also be based on the size of the chamber or the size of the product load. The
data acquired in the study are used to calculate an EtO exposure time and the
minimum process parameters expected to elicit a specific probability of survival
of the challenge organism expressed in CFUs (see "The Full Cycle,"
below). Theoretically, because all other process parameters are the same, the
statistical evaluation of the data should result in a plotted survivor curve or
regression analysis that demonstrates a consistent relationship between the EtO
gas exposure time and the number of survivors (positive BIs). Unfortunately,
this is not always the case. In addition to the issue of consistency between the
BIs, the integrity of the data is dependent on the consistent delivery of the
process parameters for each cycle, including precise control of gas injection
times that can be considered part of the gas exposure phase for the purposes of
this study. Controlling the gas injection time in large industrial EtO chambers
necessitates ensuring that the head space pressure in the gas tanks is
consistent from one cycle to another and that the gas delivery lines are
uniformly either full or purged of gas. A few minutes difference in gas inject
time can significantly change the results in this study. Other variables to
consider include the temperature of the gas volatilizer, evacuation rates and
times, gas makeups, and air exchanges.
If the survivor curve study is conducted in an industrial chamber, the data
should elicit an accurate probability of survival of a specific challenge
organism from which routine cycle parameters can be determined. The BI should be
an inoculated carrier. For example, if the BI is a spore strip that contains a
population of at least one million (106) spores of Bacillus
subtilis var. niger and the construction of the survivor curves
demonstrates that total kill was obtained at 2 hours of gas exposure time, the
full cycle used to routinely sterilize the product to an SAL of 10–6
could employ 4 hours of gas exposure. To achieve a smaller SAL, for example 10–3,
would require adding the appropriate additional gas exposure time to the
original 2 hours.
The BI testing process should be validated prior to initiation of the
survivor curve study to obtain an acceptable and documented recovery method. The
process begins with macerating the BIs in sterile water. The suspension is
evaluated by plating specific dilution aliquots of serial dilutions of the
suspension onto a selected agar medium and counting the number of CFUs after
incubation. The number and extent of dilutions needed will be based on the
duration of gas exposure and concentration in relation to the number of BI
organisms. In other words, more dilutions are required when more survivors are
expected.
A statistical analysis made from each cycle of survival data should show the
log10 of the surviving population plotted against EtO exposure time
intervals. The best-fit rectilinear curve through the data can be drawn or
determined by regression analysis using the method of least squares.8
The survivor curve method is complicated by the number of serial dilutions
that must be prepared and the quality of the dilutions being dependent on the
skill of the person performing the test as well as the precision of the
equipment used. Only trained personnel who can adequately practice aseptic
technique should conduct this test. Calibrated pipettes and dilution controls
help ensure the test's accuracy. The dilutions should also be chosen to yield
counts between 30 and 300 CFUs. It is generally assumed that numbers ranging
from 30 to 100 CFUs should be used because it is thought that higher numbers of
CFUs per plate could result in inaccurately low counts and that numbers lower
than 10 CFUs per plate could give unreliable counts. For practical purposes,
counts between 30 and 300 are generally acceptable.
Fraction-Negative Method. The
fraction-negative method also involves exposing BIs to multiple cycles of graded
exposures to EtO. The differences between the two methods are the number of
cycles recommended and the number used to enumerate the survivors (in terms of
the positive BIs).4 A minimum of seven cycles should be employed in
this study, each utilizing different gas exposure time periods. These seven
cycles should elicit the following survivor data:
- At least one sample set that elicits all survivors (growth in all BIs
tested).
- At least four sample sets that elicit fractional data, i.e., a fraction of
the BIs in each set demonstrates growth or survival.
- At least two sample sets in which there is neither growth nor survivors.
The method used to enumerate the number of positive BIs is more
straightforward than in the survivor curve method. In the fraction-negative
method, the BIs are immersed directly into the appropriate media and incubated.
Results are recorded in terms of the total number of BIs demonstrating growth
and the total number eliciting no growth for each set of test samples.
METHODS FOR DETERMINING CYCLE LETHALITY
Calculating D-Values. The D-value
corresponding to the survivor curve construction results is determined by either
reading it from a graph or calculating from the data the relevant time interval
for reducing the count by one log (Table I). The
performance of the fraction-negative method will generate data that are then
utilized to calculate a D-value using a method described by Pflug and Holcomb.11
This method is also referred to as the full or generic Spearman-Karber
procedure. It does not require the use of the same number of replicates nor the
same gas exposure time intervals because it involves spore strips rather than
counting CFUs.
| Time of Exposure to Sterilant (t) |
Number of Test Samples Exposed (n) |
Number of Test Samples Showing No Growth (r) |
| t1 |
n1 |
r1 |
| t2 |
n2 |
r2 |
| t3 |
n3 |
r3 |
| t4 |
n4 |
r4 |
| t5 |
n5 |
r5 |
| t6 |
n6 |
r6 |
| t7 |
n7 |
r7 |
Table I. Variables used in calculating D-values
during MPQ.
In Table I, t1 represents the shortest exposure time to
sterilant, and all test samples run through this short time frame should show
growth. The variables t2 to t5 correspond to
increasing exposure times in the fraction-negative area, otherwise known as the
quantal region. Exposure times t6 and t7
should represent tests in which none of the tests show growth.
In the following equations, r1 is the number of test
samples out of the number exposed (ni) that show no growth at
exposure time t1. For each period of exposure to sterilant t1
to t6, the factors x and y are calculated as
shown:
At ti, all test samples show growth, therefore,
From the calculated values of xi and yi
above, the value µI (mean time to attain no growth) can be
calculated for each period of exposure (t1 to t6)
as follows:
The mean time to attain no growth from all of the test samples, µ, can be
calculated from the sum of µI for each time of exposure t1
to t6:
Where the interval between exposure times (d) is constant and the same number
of test samples (n) is used at each exposure time, the mean to attain no
growth (µ) can be calculated from the equation:
The mean D-value (D) can be calculated from the equation:
where N0 = the initial number of challenge organisms per
test sample.
For the purposes of calculating the sterilization period using the Dcalc
method below, the upper 95% confidence level for D should be used. This can be
calculated from the equation:
where V is derived as follows:
In addition to the method described above, there are three other commonly
used means of calculating D-values. The limited Spearman-Karber procedure
requires the same number of replicates for each exposure time and equal
intervals between the various gas exposure times. For example, the exposure
times of 9, 12, 15, and 18 minutes provide an arithmetic series of 3-minute time
intervals.
The Stumbo-Murphy-Cochran procedure requires one result in the
fraction-negative range, consisting of time (t), the number of units
negative for growth (r) and the number of replicates (n) at each
exposure time as well as the initial number of microorganisms per replicate (N0).
The D-value for each exposure time is calculated from the following equation.
The confidence limits of the mean may be calculated by conventional procedures
using natural logs.
The factors in the equation represent the following:
U = exposure interval.
N0 = initial number of organisms per replicate carrier
unit.
NU = ln (n/r).
n = total number of replicate units at exposure U.
r = number of units negative for growth at U.
The D-values are averaged to obtain the overall D-value for the experiment.
The limited Stumbo-Murphy-Cochran procedure requires only one exposure time
in the fraction-negative range.
It is important to choose the most appropriate procedure of the three
described to calculate the D-value and its upper confidence limit. Certain
methods have limitations. For example, Graham and Boris have noted that a
confidence interval cannot always be determined from the Stumbo-Murphy-Cochran
procedure.12 Shintani et al. have observed that the Stumbo-Murphy-Cochran
procedure can yield invalid results if r is close to n.13
They also state, however, that this procedure may be superior in situations in
which the sample population is at least 50, r is at least 1, and r/n
is less than 0.9.
Half-Cycle Method. The half-cycle method determines the minimum time a
specific product load must be exposed to an EtO process to guarantee that no
survivors exist from the BIs used to monitor the cycle's efficacy. The cycle
chosen is based not only on the elements discussed in the companion to this
article but also on the microbiological data obtained prior to determining the
parameters for the half-cycle.1 The minimum ranges of the half-cycle
are based on this predetermined microbiological data, whereas the maximum ranges
are dictated by the product and packaging tolerances established through
functional testing.
If the half-cycle method is chosen, an additional cycle should be performed,
preferably before the first half-cycle is conducted. In some cases, other
proportionate cycles, referred to as sublethal or fractional cycles, may be
used. The gas exposure time is shortened in order to obtain survivors from the
BIs used to monitor the cycles. The process parameters in a sublethal cycle are
the same as those in the half-cycle with the exception of the gas exposure time
period. Such cycles are conducted to demonstrate that positive BIs can be
recovered and to prove the adequacy of the BI, the sterility test method, and
associated equipment and materials. The sublethal cycle is not required if
either the survivor curve construction or fraction-negative method is used
because these methods require positive sterility test results as part of the
study.
A second objective in conducting a sublethal cycle is to attempt to
demonstrate that the resistance of the product's bioburden is less than that of
the BI. This is accomplished by using actual product test samples that have been
manufactured, packaged, and handled in the same manner intended for routine
production. It is imperative that these test samples were made in accordance
with routine manufacturing procedures because the bioburden on the product must
represent what the product would normally contain. Sometimes using a number of
different product samples can yield useful information when there is concern
about not only the quantity of bioburden but also where it is located. If a
number of different samples have been sterility tested, those with the highest
bioburden should be used for validation. After the completion of the sublethal
cycle, the product samples are sterility tested. Since the objective of the test
is to demonstrate sterility of the bioburden, the data can be compared to the
positive sterility test results obtained from the BIs to demonstrate that the
resistance of the product's bioburden is less than that of the BI.
A third objective of the sublethal cycle study is to determine that a PCD
used as an external monitoring device for routine sterilization cycles yields
more positive sterility tests than an internal BI does. It has been suggested
that it is acceptable to obtain positive sterility test results from those
external PCDs exposed to the half-cycles if previous comparative resistance
studies have proven that they are as resistant or more resistant than the
internal product BI and if no positive sterility test results are obtained from
the BIs placed in internal load locations. If total kill is obtained from all
PCDs on all half-cycles, it can be assumed that the resulting validated full
cycle will provide an additional safety factor that goes beyond the minimum
requirements.
Typically, the sterilization efficacy of the first half-cycle is determined
before any subsequent cycles are conducted. If the first half-cycle experiment
elicits all negative BIs, two additional half-cycles are conducted to confirm
the data. All three half-cycles must employ the same parameters to demonstrate
reproducibility and reliability. All three half-cycles must demonstrate the
ability to elicit all negative BIs. If a SAL of 10–6 is required,
a gas exposure time that is at least double that used in the half-cycle becomes
the minimum gas exposure time used routinely in the full cycle.
THE FULL CYCLE
The data obtained from the survivor curve construction method, the
fraction-negative method, or the half-cycle method are used to design the full
cycle that will be used to routinely sterilize the product. The full cycle must
be capable of reliably demonstrating a required SAL that consists of the minimum
time to obtain all negative BIs or CFUs with an additional margin of safety.
Normally this can be expressed as 10–n where n is
the cumulative probabilities of the log minimum time to sterilize and the log
margin of safety. For example, if the half-cycle method was chosen to validate a
given sterilization cycle with a required SAL of 10–6, the
half-cycle must demonstrate the ability to ensure that there are no survivors
from BIs that have a certified population of 106. Since the BIs have
a greater resistance than the product bioburden, it can be concluded that the
time is sufficient to achieve product sterility. However, adding an equivalent
sterilization time period increases the margin of safety. The additional 6 log
reduction in the population (10–6) is theoretical and is obtained
by doubling the half-cycle gas exposure time. In other words, if the half-cycle
gas exposure time period was 2 hours, the full-cycle gas exposure time would be
a minimum of 4 hours.
The processing ranges established for the critical parameters in the full
cycle should be based on the ranges determined in the half-cycle. The minimum
process parameters in the full cycle, such as temperature, sterilant
concentration, pressure, and humidity, may also be greater than those used in
the half-cycle to ensure greater lethality with the full cycle. The first full
cycle should be included in the validation to complete the profile of the
product load. The same type and amount of monitoring should be conducted during
the first full cycle as was conducted on each of the half-cycles. Comparisons of
temperature distribution throughout the chamber and the load can justify a
determination of reliability and repeatability for the process. These data can
also be useful for determining the significance of any minor changes that are
made to the load.
REPORT ON VALIDATION AND CERTIFICATION
The documentation on the validation and certification should include but not
be limited to the following:
- References to the maintenance and calibration procedures for the
processing equipment.
- Specifications for the EtO chamber.
- References to the commissioning data.
- An indication that all gauges, monitoring devices, etc., were calibrated
prior to initiation of and at the conclusion of the validation.
- The validation protocol.
- Comparative resistance test data and reference to the protocol.
- A complete description of the products used as test samples, including
packaging.
- BI certification and all data, according to current standards, required in
the labeling.
- The physical and biological records from all of the validation cycles.
- Placement diagrams and data for the monitoring devices (including
temperature, gas concentration, and relative humidity).
- Documentation on and verification of the positioning and location where
the BIs and product test samples were placed on the product load.
- BI population determination test results and statement of retention
samples.
- All biological test data associated with the validation.
- EtO residual test data.
- References to the test procedures used during the validation.
- Documentation of operating procedures, including process control limits.
- A test report summarizing and analyzing the validation data.
ANNUAL REVALIDATION
An annual revalidation of the process should be conducted to verify the
integrity of the original validation data. It may be advisable to conduct this
revalidation within 13 months of the initial validation and each year thereafter
until sufficient data are obtained to allow that time period to be extended.
Afterwards, revalidation should be conducted at least every 2 years.
Data from each revalidation should be compared with that from the original
validation and any subsequent revalidation to confirm that the original
performance specification remains valid. A revalidation consists of a thorough
review of all the factors that could affect the sterilization process, including
changes to the product or its packaging, the manufacturing methods and facility,
and component and materials suppliers. Product bioburden test data should be
analyzed and trends noted to ensure that there have been no substantial
unacceptable changes in quantity or characterization. The testing conducted of
the preconditioning room, the sterilization chamber and ancillary equipment, and
the aeration room and equipment should be compared with the data obtained during
the original validation as well as to the commissioning data. All associated
programs, such as preventive maintenance and calibration, should be reviewed for
effectiveness.
CONCLUSION
A well-designed and scientifically sound MPQ helps ensure that the process is
safe, efficacious, and efficient. Performed well, the MPQ should deliver the
required SAL for the product in an economical process and should enable the
process to continue without having to be repeated. It should also prevent
product reprocessing and delays in product release to the marketplace. A
knowledgeable approach that considers all of the various aspects of the MPQ will
ensure the success of the process.
ACKNOWLEDGMENT
The authors would like to express their gratitude to Aubrey S. Outshoorn,
PhD, of the United States Pharmacopeial Convention Inc. for his contributions to
this article.
REFERENCES
1. PJ Sordellini and SE Satter, "EtO Sterilization: Principles of
Process Design," Medical Device & Diagnostic Industry 20, no.
12 (1998): 47–59.
2. "Sterilization of Medical Devices—Microbiological Methods—Part 1:
Estimation of Population of Microorganisms on Products," ANSI/AAMI/ISO
Standard 11737-1 (Arlington, VA:
Association
for the Advancement of Medical Instrumentation, 1995).
3. "Guidance on the Application of EN 29001 and EN 46001 and of EN 29002
and EN 46002 for Non-Active Medical Devices," BS/EN 724 (London: British
Standards Institution, 1995).
4. "Medical Devices—Validation and Routine Control of Ethylene Oxide
Sterilization," ANSI/AAMI/ISO Standard 11135 (Arlington, VA: Association
for the Advancement of Medical Instrumentation, 1994).
5. EW Nester et al., Microbiology, 2nd ed. (Chicago: Holt, Rinehart,
and Winston, 1978).
6. The
Center for
Devices and Radiological Health, FDA Guide for Validation of Biological
Indicator Incubation Time (Rockville, MD, FDA, 1985).
7. United States Pharmacopeia 23, 8th supp., monograph <71>,
1998.
8. "Sterilization of Health Care Products—Biological Indicators—Part
1: General Requirements," ANSI/AAMI/ISO Standard 11138-1 (Arlington, VA:
Association for the Advancement of Medical Instrumentation, 1994).
9. "Sterilization of Health Care Products—Biological Indicators—Part
2: Biological Indicators for Ethylene Oxide Sterilization," ANSI/AAMI/ ISO
Standard 11138-2 (Arlington, VA: Association for the Advancement of Medical
Instrumentation, 1994).
10. "Contract Sterilization for Ethylene Oxide," AAMI TIR no.
14-1997 (Arlington, VA: Association for the Advancement of Medical
Instrumentation, 1997).
11. IJ Pflug and RG Holcomb, "Principles of Thermal Destruction of
Micro-Organisms," in Disinfection, Sterilization and Preservation,
ed. SS Block, 3rd ed. (Philadelphia: Lea and Febiger, 1983).
12. GS Graham and CA Boris, "Chemical and Biological Indicators,"
in Sterilization Technology: A Practical Guide for Manufacturers and Users of
Health Care Products (New York: Van Nostrand Reinhold, 1993).
13. H Shintani et al., "Comparison of D10-Value Accuracy by
the Limited Spearman-Karber Procedure (LSKP), the Stumbo-Murphy-Cochran
Procedure (SMCP), and the Survival-Curve Method (EN)," Biomedical
Instrumentation Technology 29, no. 2 (1995): 113–124.
Susan Edel Satter of Satter & Associates Inc. (Boulder, CO) and Paul
J. Sordellini of Quality Solutions Inc. (Annandale, NJ) are consultants for the
medical device manufacturing industry.
Illustration by Brad Hamann
|